Desorption amount(Hz) | 166 | 491 | 732 | 1,134 | 2,296 | 6,006 | 12,170 | 19,584 | Average | σ |
---|---|---|---|---|---|---|---|---|---|---|
1Hz/sec Desorption temperature(℃) |
-13 | -12.3 | -13 | -12.4 | -12.4 | -13 | -11.8 | -10.9 | -12.4 | 0.73 |
付着量(Hz) | 1Hz/sec Desorption temperature(℃) |
---|---|
166 | -13 |
491 | -12.3 |
732 | -13 |
1,134 | -12.4 |
2,296 | -12.4 |
6,006 | -13 |
12,170 | -11.8 |
19,584 | -10.9 |
Average | -12.4 |
σ | 0.73 |
Main functions
Twin-TQCM Sensor Module | PSA-QM-1002 |
---|---|
Frequency | 10.278MHz(Fundamental wave) /30.833MHz(The 3rd. overtone) |
Mass sensitivity | 2.39x108 (Hz/g)㎠(Fundamental wave) 7.17x108 (Hz/g)㎠(The 3rd. overtone) |
Operating temperature range | -80 ~ +125℃ |
Temperature characteristics | ±10ppm against +25℃ (-80℃ to +80℃) |
Dimensions | Φ35.0x23.3(H)mm Electrode area:12.57㎟ |
Weight | ≦50g |
Peltier power | ≦9W |
Measurement viewing angle | 20° (Harf Angle) |
Twin-QCM Sensing unit | 4ch type・PSA-QS-1001 | 1ch type・PSA-QS-1002 |
---|---|---|
Number of control channels | 1 ~ 4 | 1 |
Frequency detection accuracy | ≦1ppm | ≦1ppm |
Frequency resolution | 0.01Hz | 0.01Hz |
Operating temperature range | +10 ~ 40℃ | +10 ~ 40℃ |
Power supply voltage | AC100V ~ AC240V(50/60Hz) | AC100V ~ AC240V(50/60Hz) |
Dimensions | 437(W) x 132(H) x 499(D) mm(Except protrusions) | 437(W) x 132(H) x 499(D) mm(Except protrusions) |
Weight | ≦10kg | ≦10kg |
Twin-QCM Sensing unit | 1ch type・PSA-QS-1002 |
---|---|
Number of control channels | 1 |
Frequency detection accuracy | ≦1ppm |
Frequency resolution | 0.01Hz |
Operating temperature range | +10 ~ 40℃ |
Power supply voltage | AC100V ~ QC240V(50/60Hz) |
Dimensions | 437(W) x 132(H) x 499(D) mm(Except protrusions) |
Weight | ≦10kg |